- Speed Your Medical Equipment to Market -






TABLE VALUES AND TEST LIMITS FOR IEC 60601-1, UL 2601-1, EN 60601-1, & CSA C22.2 No. 601.1

All Amendment changes and US Deviations have been integrated in these tables.

 

MINIMUM GAPS TO AVOID CRUSH/PINCH POINTS

Requirements for Medical Electrical Equipment (Covered by IEC 60601-1)

From EN 394:  Safety of machinery - Minimum gaps to avoid crushing of parts of the human body (1993)

Crushing Zone
Zone in which the human body or parts of the human body are exposed to a crushing/pinching hazard. 
This hazard will be generated if:
- two movable parts are moving towards one another;
- one movable part is moving towards a fixed Part

Methodology
a) Identify the crushing hazards.
b) Assess the risks from these hazards
c) From table, select the appropriate minimum gap relating to the body part at risk 
d) If adequate safety cannot be achieved by the minimum gaps selected from the table, other or additional measures and/or means shall be used

Note: A crushing zone is considered only for powered movements or where weight or momentum may generate sufficient force to generate a crushing/pinching hazard.

The possibility of access to a crushing zone for a particular part of the body is dependent on the following:
- The gap between the parts;
- The depth of the crushing zone;
- The dimensions c of the opening in the protective structure and its distance d from the crushing zone.  


 

HELPFUL CONVERSIONS

1 cm = 0.394 in.      (1 m = 39.4 in)
1 in. = 2.54 cm        (1 ft. = 30.48 cm)

10 degree tilt: (Distance One Side Lifted) = (Width of Base)(0.173648)

F  = (C  x 1.8) + 32
C  = (F  / 1.8 ) - 32

1 N = 0.225 lbf
1 kg = 2.2 lb            (1 g = 0.0022 lb)

1 m/sec = 2.237 mph

1 Nm = 141.6 in.oz = 0.7376 ft.lbs = 8.851 in.lbs

1 l = 33.8 oz    (200 ml = 6.76 oz) 
1 l = 61 in2